Expert in Water Quality Measurement and Water Treatment Project Since 2007
YLG-2058-01
INDUSTRIAL RESIDUAL CHLORINE SENSOR
Measuring Principle
Electrolyte and osmotic membrane separates the electrolytic cell and water samples, permeable membranes can selectively to ClO- penetration; between the two electrode has a fixed potential difference, current intensity generated can be converted into residual chlorine concentration;
At the cathode: ClO- + 2H+ + 2e- → Cl-+ H2O
At the anode: Cl- + Ag → AgCl + e-
Because in a certain temperature and pH conditions, HOCl, ClO- and residual chlorine between fixed conversion relationship, in this way can measure the residual chlorine.
TECHNICAL SPECIFICATIONS
Model | YLG-2058-01 Industrial Residual Chlorine Electrode |
Principle | Polarographic |
Measuring range | 0.005 ~ 20ppm(mg/L) HOCl |
The minimum detection limit | 5ppb or 0.05mg/L HOCl |
Accuracy | 2% or ±10ppb |
HOCl response time | 90%<90second |
Storage temperature | -20 ~ 60 ℃ |
Operation temperature | 0~45℃ |
Sample temperature | 0~45℃ |
Calibration method | laboratory comparison method |
Contact Us
Contact us right away
BOQU Instrument focus on development and production of water quality analyzers and sensors, including water quality meter, dissolved oxygen meter, pH sensors, etc.