Analysis of COD measuring instrument detection principle

Before the COD rapid determination instrument detects the COD content value,

the sample needs simple pretreatment.


After adding the chemical oxygen demand reagent for digestion, the detection work is carried out. The digestion process of the water sample directly

determines whether the organic matter in the sample can be completely

oxidized.


When using a chemical oxygen demand measuring instrument to measure the

chemical oxygen demand content in the determination, the chemical oxygen demand

digestion instrument needs to be heated to a constant temperature (165°C, 15

minutes), and then placed in the digestion colorimetric integrated tube.

Otherwise, the organic matter in the sample cannot be completely oxidized, and

the detection result of chemical oxygen demand is lower than the actual

value.


The principle of the COD analyzer uses a closed digestion tube for closed digestion. A special oxidant containing a certain amount of potassium dichromate is used in a strong acid solution. Under the action of a catalyst, the water sample is digested at a constant temperature of 165°C for 15 minutes to oxidize the water. The reducing substance.


From the principle of detecting chemical oxygen demand, it can be seen that

detecting chemical oxygen demand requires constant heating for 15 minutes in an the environment of 165°C.


However, if the COD digester is first put into the digestion colorimetric

integrated tube and then heated, it will take a certain time for the COD

digester to heat up from room temperature to 165°C, which will cause the COD to

be at 165°C.


The heating time in the C environment is much less than 15 minutes, which

leads to the incomplete reaction of the organic matter in the sample, which

leads to the detection of the chemical oxygen demand value far lower than the

actual value.


Therefore, after the digester is heated to the preset temperature, the

digestion colorimetric integrated tube needs to be placed.


In the routine test parameters, in addition to the cod measuring instrument

and the multi-functional intelligent digestion instrument, the total phosphorus

measuring instrument and the total nitrogen measuring instrument also need to be

used in conjunction with the digestion instrument, such as the digestion

condition of total phosphorus: 120℃, 30 minutes; the total nitrogen digestion

condition is 120°C, 40 minutes; when taking the water sample of the instrument

for testing, the digestion process needs to be heated to a constant temperature

and then put into the digestion colorimetric integrated tube.

GET IN TOUCH WITH Us
recommended articles
knowledge Project Info center
no data

Contact Us

Contact Person:Michael Luo
E-MAIL:michael@shboqu.com
Tel:86-021-20965230
Fax:86-021-20981909
Skype:+86-15000087545
Whatsapp:86-15000087545
Office Add:No. 118 Xiuyan Road,Pudong New Area,Shanghai,Zip Code:201315,China

Contact us right away

BOQU Instrument focus on development and production of water quality analyzers and sensors, including water quality meter, dissolved oxygen meter, pH sensors, etc.

Copyright © 2025 Shanghai BOQU Instrument Co.,Ltd | Sitemap
Contact us
whatsapp
contact customer service
Contact us
whatsapp
cancel
Customer service
detect