Trouble analysis and elimination of common instruments of water quality detector


Shipping problems:

(1) Sampling fault-low pressure; short sampling time; loop fault; bad sampling valve.

(2) Reagent leakage-aging of the pipeline; two three-way valves cracked.

(3) Reagents cannot enter actively-the photocoupler is out of control or check whether the photocoupler and the coupling base are well connected; check the photocoupler circuit.

(4) The peristaltic pump does not collect water-the motor drive voltage is not enough, the driver fails or the hose is deformed and aging.

Other problems:

(1) Incomplete action-debris blocking or blocking the valve core; circuit failure.

(2) The temperature of the crusher is too high or too low-the platinum resistance is damaged; the temperature controller is damaged; the heating device is damaged.

(3) The AD value is the too low or abnormal-the photometer power supply is faulty; the adjustable resistance is not well adjusted; the light-emitting diode is damaged or aging; the photodiode is damaged; the AD module is damaged.

(4) Ultra-high zero-point voltage-light gate control fails; the zero-point drift of the photometer is large.

(5) No flow-no current and voltage of the flow probe; the flow probe should have current and voltage but no signal voltage; the data interface is not well connected; the AD module is damaged (no AD signal).

Ammonia nitrogen online monitor (gas-sensing electrode method):

(1) High measured value-inaccurate or invalid calibration fluid; air-permeable membrane with bubbles; dirty air-permeable membrane; electrode failure; aging or damage of the air-permeable membrane.

(2) Low measurement value-inaccurate or invalid calibration solution; lack of reagent; slow electrode response; pollution of air-permeable membrane; electrode failure; aging of the air-permeable membrane.

(3) Calibration is invalid-the calibration fluid is not accurate or invalid; the calibration fluid is lacking; the electrode responds slowly; the air-permeable membrane is fouled; the electrode is invalid; the air-permeable membrane is aging.

(4) The pool temperature is abnormal and the temperature sensor is faulty, and the ambient temperature exceeds the ambient temperature scale required by the instrument.

GET IN TOUCH WITH Us
recommended articles
knowledge Project Info center
no data

Contact Us

Contact Person:Michael Luo
E-MAIL:michael@shboqu.com
Tel:86-021-20965230
Fax:86-021-20981909
Skype:+86-15000087545
Whatsapp:86-15000087545
Office Add:No. 118 Xiuyan Road,Pudong New Area,Shanghai,Zip Code:201315,China

Contact us right away

BOQU Instrument focus on development and production of water quality analyzers and sensors, including water quality meter, dissolved oxygen meter, pH sensors, etc.

Copyright © 2025 Shanghai BOQU Instrument Co.,Ltd | Sitemap
Contact us
whatsapp
contact customer service
Contact us
whatsapp
cancel
Customer service
detect