Calibration of CODS-3000-01 Online UV COD Sensor
Step 1: Wire connection
Brown:+12VDC
Black:AGND
Blue:RS485 A
White:RS485 B
Yellow + Green:Ground Lead
Step 2:Open test software
Choose right Port and click “open”
Click “UVCOD” and input Sensor address and Baud rate.
Step 3: Set COD factor before calibration
1)Factor must be as 1 before calibration,
2)If not 1, please set it to 1 in configuration area.
3)Then press “确定”(OK)
Step 4: Set Wiper
1) Wiping cycle input 4096, then click”确定”,
2) Then the wiper will run in every test time.
3) Note:”确定” means “OK”; “取消” means “Cancel”
Step 5: Select"Water sample"
1)Select “water sample” in Curve type,
2)Then press “确定”(OK)
Step 6:
1)Prepare a 2L beaker,
2)Put buffer in it and the COD buffer value is 200mg/L in this test video.
3)Shake the sensor and start the wiper
4)Ensure there are no air bubbles on the sensor measuring membrane
5) Waitting 5 minutes to value stable
Step 7:
1)After value stable, please write down the UV254 value, here is 212.41
2)Input this value in Water sample 2[m-1]: 212.41;
3)After input, press “Enter” and click “确定”(OK).
4)Then input buffer value:200 in Water Sample 2[mg/L]:200
5)After input, press “Enter” and click “确定”(OK).
Step 8:
1)Then there is calibration recording in status display area.
2)Final step: change to Wiping cycle back to normal as need
3)Calibration finished.